Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Hip Preservation System Signature Series PreBent Burs under the following Recalled by ConMed Corporation Due to Metal shavings released from burs during use are...

Date: December 21, 2015
Company: ConMed Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact ConMed Corporation directly.

Affected Products

Hip Preservation System Signature Series PreBent Burs under the following labels: 1) PreBent Spherical Bur, 5.5mm x 19cm Catalog Number HPS-HB11, 2) PreBent Polishing Bur, 5.5mm x 19cm Catalog Number HPS-HB12 and 3) PreBent Oval Bur, 6.0mm x 19cm Catalog Number HPS-HB13 The ConMed Hip Preservation System Signature Series is intended to be used for resection of osseous tissue (bone) in arthroscopic and endoscopic surgical procedures

Quantity: 991 pieces

Why Was This Recalled?

Metal shavings released from burs during use are due to contact between the bur and the hood of the PreBent Bur assembly. During the manufacturing process an assembly error occurred which caused inadvertent mixing of two sizes of bearings.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About ConMed Corporation

ConMed Corporation has 69 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report