Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

REX (Recovery Exercise X-Trainer) Pressure Therapy System- USER Manual. Intended Recalled by Compass Health Brands (Corporate Office) Due to Error in the Introduction section of the user...

Date: March 7, 2023
Company: Compass Health Brands (Corporate Office)
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Compass Health Brands (Corporate Office) directly.

Affected Products

REX (Recovery Exercise X-Trainer) Pressure Therapy System- USER Manual. Intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas. Models: DVTREX-U and DVTREX -L

Quantity: 30 units

Why Was This Recalled?

Error in the Introduction section of the user manual that indicates the device may be used for the prevention of deep vein thrombosis (DVT). Marketed without a 510k for this indication

Where Was This Sold?

This product was distributed nationwide across all 50 states.