Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

YXLON Cheetah cabinet X-Ray System Product Usage: The products are Recalled by Comet Technologies Usa Inc Due to Failure to meet the requirement of the performance...

Date: October 1, 2018
Company: Comet Technologies Usa Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Comet Technologies Usa Inc directly.

Affected Products

YXLON Cheetah cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and quality control

Quantity: 77

Why Was This Recalled?

Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected