Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

115 V Blanketrol II Recalled by Cincinnati Sub-Zero Products Inc Due to The incorrect voltage component was placed into the...

Date: January 8, 2014
Company: Cincinnati Sub-Zero Products Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Cincinnati Sub-Zero Products Inc directly.

Affected Products

115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165.

Quantity: 7 units

Why Was This Recalled?

The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but

Where Was This Sold?

This product was distributed to 1 state: MI

Affected (1 state)Not affected

About Cincinnati Sub-Zero Products Inc

Cincinnati Sub-Zero Products Inc has 6 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report