Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea Recalled by Childrens Medical Ventures Due to The battery pack wire harness is improperly assembled...

Date: April 29, 2014
Company: Childrens Medical Ventures
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Childrens Medical Ventures directly.

Affected Products

Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.

Quantity: 112

Why Was This Recalled?

The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.

Where Was This Sold?

This product was distributed to 14 states: AZ, FL, IN, KS, MA, MI, MN, NV, SC, TX, UT, VA, WV, WI

Affected (14 states)Not affected

About Childrens Medical Ventures

Childrens Medical Ventures has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report