Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Spirit Plus Bed Recalled by CHG Hospital Beds Inc Due to Stryker Medical is initiating a voluntary recall of...

Date: June 28, 2016
Company: CHG Hospital Beds Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact CHG Hospital Beds Inc directly.

Affected Products

Spirit Plus Bed, A-C Powered Hospital Bed

Quantity: 4,308 units total

Why Was This Recalled?

Stryker Medical is initiating a voluntary recall of the Spirit Select and Spirit Plus A-C powered hospital beds due to reports of hi-lo actuators broken at the mount ends, which could cause the bed to unexpectedly lower resulting in patient injury.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About CHG Hospital Beds Inc

CHG Hospital Beds Inc has 3 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report