Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

HLD Systems 600 Series Washer/Pasteurizer Recalled by Cenorin, LLC Due to There is a supplier quality issue with the...

Date: April 9, 2018
Company: Cenorin, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Cenorin, LLC directly.

Affected Products

HLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160; Model 610HT, 115V & Model 610HT, 230V Product Usage: Clean and thermally disinfect medical devices using full immersion pasteurization

Quantity: 14 units

Why Was This Recalled?

There is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.

Where Was This Sold?

This product was distributed nationwide across all 50 states.