Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

HLD System Recalled by Cenorin, LLC Due to The temperature sensor/control system in the HLD Systems...

Date: January 24, 2013
Company: Cenorin, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Cenorin, LLC directly.

Affected Products

HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.

Quantity: 25 HLD systems

Why Was This Recalled?

The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Cenorin, LLC

Cenorin, LLC has 6 total recalls tracked by RecallDetector.