Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Celltrion DiaTrust COVID-19 Ag Rapid Test Recalled by CELLTRION USA INC Due to The affected lot was placed under quarantine in...

Date: December 28, 2021
Company: CELLTRION USA INC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact CELLTRION USA INC directly.

Affected Products

Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02

Quantity: 243 kits (6,075 tests)

Why Was This Recalled?

The affected lot was placed under quarantine in September 2021 due to complaints of alleged false positives, but 243 kits were erroneously released to customers in December 2021.

Where Was This Sold?

This product was distributed to 6 states: CA, FL, NJ, NY, PA, SC

Affected (6 states)Not affected

About CELLTRION USA INC

CELLTRION USA INC has 5 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report