Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in Recalled by Carestream Health, Inc. Due to After pressing and releasing the Z-Axis Motorized buttons...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Carestream Health, Inc. directly.
Affected Products
DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. CATALOG #(s): 8624181 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8624199 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F 8621617 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8622359 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F Note: PDU is a standalone Power Distribution Unit (PDU), plugged into the hospital outlet, which supplies Alternate Current (AC) to the Overhead Tube Crane (OTC).
Quantity: 72 units
Why Was This Recalled?
After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Carestream Health, Inc.
Carestream Health, Inc. has 17 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report