Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BD Pyxis MedStation ES (Med ES Main) Recalled by CareFusion 303, Inc. Due to Drawer and/or door failures that may lead to...

Date: February 14, 2023
Company: CareFusion 303, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact CareFusion 303, Inc. directly.

Affected Products

BD Pyxis MedStation ES (Med ES Main), REF 323

Quantity: 94466 units

Why Was This Recalled?

Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out,

Where Was This Sold?

Worldwide

About CareFusion 303, Inc.

CareFusion 303, Inc. has 163 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report