Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BD Pyxis: MedStation ES (Main) Recalled by CareFusion 303, Inc. Due to If automated dispensing cabinets have specific software versions,...

Date: May 2, 2024
Company: CareFusion 303, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact CareFusion 303, Inc. directly.

Affected Products

BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327; CII Safe ES, REF: 1116-00; CII Safe ES, Desktop PC, Wired Scanner, REF: 107-254-01; CII Safe ES, Desktop PC, Wireless Scanner, REF: 107-255-01; CII Safe ES, Desktop PC, Wired Scanner 1.7.3, REF: 107-256-01; CII Safe ES, Desktop PC, Wireless Scanner 1.7.3, REF: 107-257-01

Quantity: 326

Why Was This Recalled?

If automated dispensing cabinets have specific software versions, and Component Manager was configured to "installed mode", and software patch KB 5033688 was installed in the production environment, then this will will lead to an error that prevents the user from accessing the dispensing software application, which may delay accessing medication.

Where Was This Sold?

This product was distributed nationwide across all 50 states.