Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BD Pyxis MedBank Cabinet Software v2.1.2.9 Legacy MedBank Cabinets with Recalled by CareFusion 303, Inc. Due to The automated dispensing cabinet software update included a...

Date: May 24, 2022
Company: CareFusion 303, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact CareFusion 303, Inc. directly.

Affected Products

BD Pyxis MedBank Cabinet Software v2.1.2.9 Legacy MedBank Cabinets with Matrix Drawers include: . CUBEX MedFlex operating with Windows 7 . MedBank 200 Main operating with Windows 7 . MedBank 500 Main operating with Windows 7. An automated dispensing cabinet intended to securely store and dispense medications to health care providers in a clinical setting.

Quantity: 686 cabinets

Why Was This Recalled?

The automated dispensing cabinet software update included a .dll file that is not compatible with MedBank Cabinet stations with MedCart Matrix drawers. The cabinet system will stop functioning (Matrix drawers will stop opening in the MedBank Application) leading to delay in access to medications or supplies needed in residential or procedural care facilities.

Where Was This Sold?

This product was distributed nationwide across all 50 states.