Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

BD Alaris Infusion Pump Module Model 8100 Pump Module Keypad Recalled by CareFusion 303, Inc. Due to Pump Module keypad lifting, and Fluid ingress could...

Date: March 3, 2021
Company: CareFusion 303, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact CareFusion 303, Inc. directly.

Affected Products

BD Alaris Infusion Pump Module Model 8100 Pump Module Keypad; and Pump Module Door Assembly Replacement Kit with Pump Module Keypad, Parts: 49000239; 49000346; 49000438; 49000439.

Quantity: 160300

Why Was This Recalled?

Pump Module keypad lifting, and Fluid ingress could result in 1) Unresponsive keys: module continues infusion, PC unit will not alarm and must be used for programming changes, may necessitate a different Pump Module. May result in infusion start delay/inability to titrate medication. 2) Stuck keys: PC unit alarms, module exhibits Channel Error, may result in infusion interruption/start delay

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About CareFusion 303, Inc.

CareFusion 303, Inc. has 163 total recalls tracked by RecallDetector.

Related Recalls