Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Alaris System PC Unit Model 8015. modular infusion pump and monitoring system Recalled by CareFusion 303, Inc. Due to If one or more screws or washers are...

Date: June 30, 2020
Company: CareFusion 303, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact CareFusion 303, Inc. directly.

Affected Products

Alaris System PC Unit Model 8015. modular infusion pump and monitoring system

Quantity: 824277

Why Was This Recalled?

If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.

Where Was This Sold?

This product was distributed nationwide across all 50 states.