Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Alaris Syringe Module Model 8110 Recalled by CareFusion 303, Inc. Due to Cleaning agents may damage Inter-Unit Interface (IUI) Connectors,...

Date: June 30, 2020
Company: CareFusion 303, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact CareFusion 303, Inc. directly.

Affected Products

Alaris Syringe Module Model 8110, a modular infusion pump and monitoring system

Quantity: 122423

Why Was This Recalled?

Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About CareFusion 303, Inc.

CareFusion 303, Inc. has 163 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report