Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Alaris Pump Module model 8100 manufactured between November 2011 and March 2012 Recalled by CareFusion 303, Inc. Due to The recalling firm has received reports of increased...

Date: August 9, 2017
Company: CareFusion 303, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact CareFusion 303, Inc. directly.

Affected Products

Alaris Pump Module model 8100 manufactured between November 2011 and March 2012; Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N) 10942012 between November 2011 and March 2012; Alaris Pump module Bezel Kit Assembly (P/N) 10964559) shipped between November 2011 and March 2012. The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using the IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural.

Quantity: 31,622 units (28,224 in US)

Why Was This Recalled?

The recalling firm has received reports of increased or decreased flows that have occurred in certain pumps.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About CareFusion 303, Inc.

CareFusion 303, Inc. has 163 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report