Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Alaris PC Unit 8015 utilizing 802.11 b/g or 802.11 a/b/g wireless network cards Recalled by CareFusion 303, Inc. Due to Infusion pump PCs with specific software/network cards/IP addresses...

Date: January 24, 2023
Company: CareFusion 303, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact CareFusion 303, Inc. directly.

Affected Products

Alaris PC Unit 8015 utilizing 802.11 b/g or 802.11 a/b/g wireless network cards, with IP addresses from x.x.x.224 to x.x.x.255

Quantity: 83,185

Why Was This Recalled?

Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Users with interfaces to electronic medication record or admin discharge transfer, may be unable to receive/send data; infusions may need to be manually programmed

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (50 states)Not affected

About CareFusion 303, Inc.

CareFusion 303, Inc. has 163 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report