Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

DIAMOND-FLEX CIRCULAR RETRACTOR Recalled by Carefusion 2200 Inc Due to BD, formerly CareFusion, has identified a potential risk...

Date: July 1, 2016
Company: Carefusion 2200 Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Carefusion 2200 Inc directly.

Affected Products

DIAMOND-FLEX CIRCULAR RETRACTOR, ANGLED, 40MM 5MM, Product Code 89-6114, Lot Code(s): 878971; 879366; 879404, and Date Code: C16. Designed to retract or elevate organs and tissue to provide better visualization access.

Quantity: 10

Why Was This Recalled?

BD, formerly CareFusion, has identified a potential risk associated with a weld failure which could result in the wire protruding thru the tip of the instrument when articulated. If this failure were to occur while in use in a procedure it has the potential to damages tissue or organs

Where Was This Sold?

This product was distributed to 2 states: MD, WA

Affected (2 states)Not affected

About Carefusion 2200 Inc

Carefusion 2200 Inc has 112 total recalls tracked by RecallDetector.

Related Recalls