Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

CareFusion V. Mueller CUSHING MICRO BIPOLAR FORCEPS 0.7MM TIP Recalled by Carefusion 2200 Inc Due to BD has determined that the Instructions For Use...

Date: November 2, 2022
Company: Carefusion 2200 Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Carefusion 2200 Inc directly.

Affected Products

CareFusion V. Mueller CUSHING MICRO BIPOLAR FORCEPS 0.7MM TIP, OVERALL LENGTH 6" (150MM), REF F-52000

Quantity: 239 units

Why Was This Recalled?

BD has determined that the Instructions For Use (IFU) provided by BD is missing certain content that is contained in the manufacturer's IFU. This missing content is related to the interface with the device power supply, as well as cleaning and maintenance instructions. Refer to the attached V. Mueller" Bipolar Forcep IFU.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Carefusion 2200 Inc

Carefusion 2200 Inc has 112 total recalls tracked by RecallDetector.

Related Recalls