Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Gomco Style Circumcision Clamp Trays is an instrument used to Recalled by Cardinal Health Due to Two lots of Gomco Style Circumcision Clamp Trays,...

Date: February 29, 2016
Company: Cardinal Health
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Cardinal Health directly.

Affected Products

Gomco Style Circumcision Clamp Trays is an instrument used to compress the foreskin of the penis during circumcision of a male infant.

Quantity: SSI-0034: 380 units and 10-1100: 60 units

Why Was This Recalled?

Two lots of Gomco Style Circumcision Clamp Trays, SSI-0034 and 10-1100 contained a 1.3cm circumcision clamp. They both were labelled as containing a 1.1cm clamp.

Where Was This Sold?

This product was distributed nationwide across all 50 states.