Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Genius 2 Tympanic Thermometer (discontinued) Item Code: 303000 Recalled by Cardinal Health 200, LLC Due to The frequency of calibration for the Genius Tympanic...

Date: October 28, 2019
Company: Cardinal Health 200, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Cardinal Health 200, LLC directly.

Affected Products

Genius 2 Tympanic Thermometer (discontinued) Item Code: 303000

Quantity: 30142 US; OUS 113841

Why Was This Recalled?

The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Cardinal Health 200, LLC

Cardinal Health 200, LLC has 275 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report