Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Captus 4000e Thyroid Uptake System Recalled by Capintec Inc Due to It is possible that the collimators have screw...

Date: March 3, 2020
Company: Capintec Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Capintec Inc directly.

Affected Products

Captus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430-30152 - Product Usage: is intended to be used by trained nuclear Medicine Technologists or Nuclear Medicine Physicians to perform Thyroid Uptake Procedures.

Quantity: 48 (481 additional units added 3/26/20)

Why Was This Recalled?

It is possible that the collimators have screw hole locations that were insufficiently tapped. There is a potential for the collimator assembly to fall and come into contact with a patient or operator.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Capintec Inc

Capintec Inc has 4 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report