Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The device is an X-ray system that is capable of Recalled by Canon Medical System, USA, INC. Due to VAREX changed the manufacturing of the internal board,...

Date: November 15, 2023
Company: Canon Medical System, USA, INC.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Canon Medical System, USA, INC. directly.

Affected Products

The device is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.

Quantity: 1

Why Was This Recalled?

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

Where Was This Sold?

This product was distributed to 2 states: CA, LA

Affected (2 states)Not affected

About Canon Medical System, USA, INC.

Canon Medical System, USA, INC. has 49 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report