Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Recalled by Canon Medical System, USA, INC. Due to software malfunction; It was found when a user...

Date: October 16, 2018
Company: Canon Medical System, USA, INC.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Canon Medical System, USA, INC. directly.

Affected Products

Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Product Usage: The RADREX-I is a general purpose x-ray system that employs Solid State Imager(s), SSXI, which converts x-rays directly into electrical signals which can, after appropriate processing be displayed on LCD monitors or printed to a medical grade image printer. The system console is PC based devise that allows for worklist management, image storage, image processing, image exporting and image printing. The system may be equipped with a table and/or vertical wall unit, is configurable with up to two x-ray tubes, and has an auto switching function. This system is intended for use in conjunction with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck and limbs. The system is used for image acquisition, image display and transmission/output or images to external devices.

Quantity: 46

Why Was This Recalled?

software malfunction; It was found when a user performs radiography using the wireless flat panel detector (FPD), a message window displays on the monitor stating image transmission was not completed and there was no image available. It also showed the "OK" button to reacquire image data from the FPD, and the "Cancel" button to cancel the acquisition . When the user selects the "OK" button, the same message window appears. This prompted the user to repeat the same operation several times and finally select the "Cancel" button to quit the reacquistion mode.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Canon Medical System, USA, INC.

Canon Medical System, USA, INC. has 49 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report