Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: Recalled by Canon Medical System, USA, INC. Due to Table tilts without operator involvement as a result...

Date: February 8, 2019
Company: Canon Medical System, USA, INC.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Canon Medical System, USA, INC. directly.

Affected Products

Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.

Quantity: 9 systems

Why Was This Recalled?

Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.

Where Was This Sold?

This product was distributed nationwide across all 50 states.