Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Canon Aplio ultrasound system Recalled by Canon Medical System, USA, INC. Due to After completion, ultrasound system's Stress Echo may not...

Date: September 29, 2022
Company: Canon Medical System, USA, INC.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Canon Medical System, USA, INC. directly.

Affected Products

Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550, Aplio a450 CUS-AA450

Quantity: 65

Why Was This Recalled?

After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.

Where Was This Sold?

This product was distributed to 23 states: AR, CA, CO, FL, GA, IL, IA, KS, KY, LA, ME, MI, MN, MO, MT, NY, NC, OK, PA, SC, TN, WA, WV

Affected (23 states)Not affected

About Canon Medical System, USA, INC.

Canon Medical System, USA, INC. has 49 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report