Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Canister HFC-134a /980g Recalled by Candela Corporation Due to The canister neck bushing may become loose and...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Candela Corporation directly.
Affected Products
Canister HFC-134a /980g, ALUM, 15 pack Product Number: 1600-00-0224 Revision .0C Coolant canister for use with surgical laser.
Quantity: in total 7,560 canisters (3240 additional as of 2/5/20)
Why Was This Recalled?
The canister neck bushing may become loose and separate from the body of the canister during use. The issue occurs only when the canisters are used in Candela legacy devices that utilize screw-in or threaded interface type DCD modules
Where Was This Sold?
This product was distributed to 29 states: AZ, CA, CO, FL, GA, IL, IN, IA, KS, LA, MD, MA, MI, MN, NV, NY, NC, OH, OR, PA, SC, SD, TN, TX, VT, VA, WA, WI, WY
About Candela Corporation
Candela Corporation has 5 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report