Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

C2 CryoBalloon Ablation System - Controller Cap. C2 CryoBalloon Controller Recalled by C2 Therapeutics, Inc. Due to The Controller Cap of the C2 CyroBalloon Ablation...

Date: January 6, 2017
Company: C2 Therapeutics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact C2 Therapeutics, Inc. directly.

Affected Products

C2 CryoBalloon Ablation System - Controller Cap. C2 CryoBalloon Controller; Cryosurgical Unit, Cryogenic Surgical Device The affected product is the Controller Cap component only, which is provided to the customer already assembled onto the Controller. Model number: FG-1012 Product Usage: The C2 CryoBalloon Ablation System is intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett s Esophagus with dysplasia. It is a cryosurgical system with a nitrous oxide cooled balloon that is compatible with a commercially available endoscope with a minimum working channel inner diameter (ID) of 3.7 mm and length of 100 cm. The C2 CryoBalloon System is a system comprised of a Catheter (sterile), Controller (non-sterile), and Cartridge (non-sterile). The Device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the inflated balloon, and the treatment site is selected by adjusting the endoscope and Controller position. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Controller.

Quantity: 755 controller caps

Why Was This Recalled?

The Controller Cap of the C2 CyroBalloon Ablation System may crack as it is tightened onto the controller. That could compromise the integrity and result in damage to the Controller Cap.

Where Was This Sold?

This product was distributed nationwide across all 50 states.