Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Marodyne-LiV Recalled by Btt Health Gmbh Due to Some power supply units in Austria and Germany...

Date: August 1, 2024
Company: Btt Health Gmbh
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Btt Health Gmbh directly.

Affected Products

Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment

Quantity: 62 units

Why Was This Recalled?

Some power supply units in Austria and Germany had a voltage drop in the millisecond range, which led to a cancellation of the treatment and subsequent restart of the medical device. The treatment is only restarted by a step-off/step-on (step-on for activation).

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Btt Health Gmbh

Btt Health Gmbh has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report