Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
BREG Sterile Polar Pads Recalled by Breg Inc Due to The recall was initiated because Breg has determined...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Breg Inc directly.
Affected Products
BREG Sterile Polar Pads; 02356 Rev M, PAD M/U XL STER, Mfg.Date: 112012. Model #s: 02510, 09901, 02330, 02350, 02490, 02356, 02496, 02410, 02430. Local anesthetic effect.
Quantity: 44,883 units
Why Was This Recalled?
The recall was initiated because Breg has determined that some Sterile Polar Pads products manufactured from January 2010 to October 2012 may have sustained damage to the product packaging which may compromise product sterility assurance.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Breg Inc
Breg Inc has 13 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report