Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Outlook Safety Infusion System Burette Set. For intravenous fluid administration. Recalled by B. Braun Medical, Inc. Due to Reports were received of an incorrectly assembled Outlook...

Date: August 5, 2013
Company: B. Braun Medical, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact B. Braun Medical, Inc. directly.

Affected Products

Outlook Safety Infusion System Burette Set. For intravenous fluid administration.

Quantity: 900

Why Was This Recalled?

Reports were received of an incorrectly assembled Outlook Safety Infusion System Burette Set (Product Code 375038), and the investigation determined that one of the ULTRASITE injection sites and corresponding low pressure backcheck valve was assembled upside down.

Where Was This Sold?

This product was distributed to 4 states: FL, NJ, NY, TX

Affected (4 states)Not affected

About B. Braun Medical, Inc.

B. Braun Medical, Inc. has 216 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report