Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.
Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Recalled by B. Braun Medical, Inc. Due to In affected products, the tubing outer diameter may...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact B. Braun Medical, Inc. directly.
Affected Products
Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 362033 UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; 362034 Priming volume: 19mL, length: 120in, SafeLine; 362035 INFUSOMAT SPACE PUMP IV SET; 362043 Priming volume: 39mL, length: 120in, SafeLine; 362050 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; 362420 SPACE PUMP SET 15 DROP W/2 SAFEDAY; 362432 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; 362911 Priming volume: 14mL, length: 84in, CARESITE, Spin-Lock; 363010 Priming volume: 21mL, length: 121in, SafeLine; 363019 Priming volume: 40mL, length: 123in, SafeLine; 363030 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 363032 20 drops/mL, priming volume: 20mL, length: 119in, SafeLine; 363230 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; 363410 15 drops/mL, priming volume: 23mL, length: 119in, CARESITE; 363419 Priming volume: 41mL, length: 120in, CARESITE; 363420 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; 363421 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 363422 Priming volume: 21mL, length: 130in, CARESITE; 363423 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; 363424 UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; 363430 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; 363433 UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 363901 60 drops/mL, priming volume: 11mL, length: 129in; 363902 Priming volume: 25mL, length: 130in; 363904 UNIV. 15 DROP PUMP SET, ASV; 470119 15 drops/mL, priming volume: 23mL, length: 119in, CARESITE; 480254 SPACE PUMP IV SET 15D, 110 IN.; 480255 SPACE PUMP IV SET, 127 IN; 480257 60 DROP NITRO SPACE SET, 88 IN.; 480258 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 480263 SPACE PUMP IV SET 15D, 90 IN.; 480264 SPACE PUMP IV SET, 15 DRPS/ML, 95 IN.; 490037 Priming volume: 25mL, length: 136in; 490038 60 drops/mL, priming volume: 11mL, length: 129in; 490100 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 490101 Priming volume: 19mL, length: 120in, CARESITE; 490102 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; 490103 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; 490104 60 drops/mL, priming volume: 21mL, length: 115in, CARESITE; 490105 Priming volume: 39mL, length: 120in, CARESITE
Quantity: 33,820,976 units
Why Was This Recalled?
In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About B. Braun Medical, Inc.
B. Braun Medical, Inc. has 216 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report