Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Diacap(R) Ultra Dialysis Fluid Filter Recalled by B. Braun Medical, Inc. Due to There is the potential of the residual moisture...

Date: March 1, 2013
Company: B. Braun Medical, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact B. Braun Medical, Inc. directly.

Affected Products

Diacap(R) Ultra Dialysis Fluid Filter, Catalog Number 7107367. Intended to filter bacteria and endotoxins from dialysis fluid.

Quantity: 1580

Why Was This Recalled?

There is the potential of the residual moisture of the Diacap Ultra membrane being reduced over time, which results in the reduction of the membrane permeability (ultrafiltration coefficient).

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About B. Braun Medical, Inc.

B. Braun Medical, Inc. has 216 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report