Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Recalled by B Braun Medical Inc Due to Firm learned of the potential for cracks to...

Date: January 28, 2022
Company: B Braun Medical Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact B Braun Medical Inc directly.

Affected Products

Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number 3456103A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI 04046964285608), 710200U (UDI 04046964285615), 710200K (UDI 04046963686338), 710500L (UDI Not applicable) and 710500K (UDI Not applicable).

Quantity: 95

Why Was This Recalled?

Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About B Braun Medical Inc

B Braun Medical Inc has 148 total recalls tracked by RecallDetector.

Related Recalls