Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
CARESITE SMALLBORE EXT SET T-PORT 5 IN. Recalled by B. Braun Medical, Inc. Due to The extension set has a high probability of...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact B. Braun Medical, Inc. directly.
Affected Products
CARESITE SMALLBORE EXT SET T-PORT 5 IN., Product Code 470114, 100 units per carton. The A6 Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood.
Quantity: 354,200 US; 1,800 OUS
Why Was This Recalled?
The extension set has a high probability of potential failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About B. Braun Medical, Inc.
B. Braun Medical, Inc. has 216 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report