Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brand Name: Perfusor Space Product Name: PERFUSOR SPACE ACM-KIT Model/Catalog Recalled by B Braun Medical Inc Due to Perfusor Space Syringe Pump's instructions for use previously...

Date: June 14, 2024
Company: B Braun Medical Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact B Braun Medical Inc directly.

Affected Products

Brand Name: Perfusor Space Product Name: PERFUSOR SPACE ACM-KIT Model/Catalog Number: 8713040U Product Description: Perfusor Space, Infusion Pump Kit, 1 each Component: N/A

Quantity: 930

Why Was This Recalled?

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About B Braun Medical Inc

B Braun Medical Inc has 148 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report