Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ExacTrac Dynamic software Recalled by Brainlab AG Due to Display of potential patient movement might be delayed...

Date: May 14, 2021
Company: Brainlab AG
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Brainlab AG directly.

Affected Products

ExacTrac Dynamic software, Model 20910-01B ETD Positioning and Monitoring Installer 1.0.2 - Product Usage: intended to position patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, to monitor the patient position, and to provide a beam hold signal in case of a deviations in order to treat lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Quantity: 22 systems in the U.S.

Why Was This Recalled?

Display of potential patient movement might be delayed to the user for high dose treatments.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Brainlab AG

Brainlab AG has 39 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report