Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Axial Torque Limiting Driver (1st generation) Product Usage: The Bradshaw Recalled by Bradshaw Medical Inc. Due to A component failure could cause a situation where...

Date: February 2, 2015
Company: Bradshaw Medical Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Bradshaw Medical Inc. directly.

Affected Products

Axial Torque Limiting Driver (1st generation) Product Usage: The Bradshaw Medical Axial Torque Limiting Driver is a reusable surgical instrument designed to limit the magnitude of applied (input) torque transferred to an external (output) device or fastener in the clockwise direction. The device is a non-adjustable mechanical click style hand torque tool which presents an audible and tactile click when the maximum desired torque is reached.

Quantity: 3,760 units

Why Was This Recalled?

A component failure could cause a situation where the applied torque was significantly higher than the calibrated torque setting.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Bradshaw Medical Inc.

Bradshaw Medical Inc. has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report