Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Driver Instrument Recalled by Boston Scientific Neuromodulation Corporation Due to Pending update to indirect decompression system instructions for...

Date: May 12, 2023
Company: Boston Scientific Neuromodulation Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Boston Scientific Neuromodulation Corporation directly.

Affected Products

Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Technique Manual

Quantity: 21,533

Why Was This Recalled?

Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.