Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Torflex Transseptal Guiding Sheath Kit (TFK-M) TF85-32-63-45 TORFLEX 8.5F 63CM Recalled by Boston Scientific Corporation Due to Potential for hole in the Tyvek layer of...

Date: June 10, 2025
Company: Boston Scientific Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Boston Scientific Corporation directly.

Affected Products

Torflex Transseptal Guiding Sheath Kit (TFK-M) TF85-32-63-45 TORFLEX 8.5F 63CM 45DEG TF85-32-63-55 TORFLEX 8.5F 63CM 55DEG Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation / puncture.

Quantity: 14,724

Why Was This Recalled?

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Boston Scientific Corporation

Boston Scientific Corporation has 601 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report