iLab Ultrasound Imaging System Recalled by Boston Scientific Corporation Due to Specific iLab Ultrasound Imaging Systems using newest 3.0...

Date: April 3, 2014
Company: Boston Scientific Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Boston Scientific Corporation directly.

Affected Products

iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology.

Quantity: 9 units

Why Was This Recalled?

Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.

Where Was This Sold?

This product was distributed nationwide across all 50 states.