Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Boston Scientific Recalled by Boston Scientific Corporation Due to Boston Scientific is voluntarily implementing a Medical Device...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Boston Scientific Corporation directly.
Affected Products
Boston Scientific, PT Graphix Magnet J-TIP Guidewire with ICE Hydrophilic Coating, Sterile EO, UPN H7491490201MJ2 (Carton Label) ,H7491490201MJo (UPN referenced on the Pouch Label) . Boston Scientific PT Graphix Magnet Guidewire with ICE Hydrophilic Coating is a steerable guidewire available with a nominal diameter of 0.014 in (0.37 mm) and in nominal lengths of 182 cm. PT Graphix Magnet is available with a shapeable Straight Tip or a preformed "J"¿ Tip. The PT Graphix Magnet Guidewire features a 10 or 38 cm polymer sleeve, coated with ICE Hydrophilic Coating, jacketing the distal tapered core wire. The proximal section of the PT Graphix Magnet Guidewire is coated with a fluorinated polymer.
Quantity: 47 units
Why Was This Recalled?
Boston Scientific is voluntarily implementing a Medical Device Recall of one batch of PT Graphix Magnet J-TIP Guidewires. Boston Scientific has found that, while the batch is labeled as a J-tip model, the guidewires are straight and do not have the pre-formed J-tip. The most severe injury expected to occur is a limited procedure delay caused by the need to replace the guidewire with one that has
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Boston Scientific Corporation
Boston Scientific Corporation has 601 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report