Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Boston Scientific Fetch 2 Aspiration Catheter Recalled by Boston Scientific Corporation Due to Boston Scientific is recalling all models of Fetch...

Date: March 22, 2016
Company: Boston Scientific Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Boston Scientific Corporation directly.

Affected Products

Boston Scientific Fetch 2 Aspiration Catheter, Catalog No. 109400-001. Product The Fetch 2 Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the peripheral and coronary vasculature.

Quantity: 21,155 units

Why Was This Recalled?

Boston Scientific is recalling all models of Fetch 2 Aspiration Catheter as a result of receiving reports from the field of catheter shaft breaks; including cases requiring the use of a snare or other intervention to retrieve the http://webapps.cdrh.fda.gov/division-tracking/images/trash.pngbroken shaft from the patient.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Boston Scientific Corporation

Boston Scientific Corporation has 601 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report