Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Boston Scientific Recalled by Boston Scientific Corporation Due to Boston Scientific is voluntarily recalling its Chariot" Guiding...

Date: November 19, 2015
Company: Boston Scientific Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Boston Scientific Corporation directly.

Affected Products

Boston Scientific, Chariot ST-CCV Guiding Sheath, Sterilized using ethylene oxide: Product Description Material Number (UPN) Guiding Sheath 5F, 45 cm, ST, CC H74939277545110 Guiding Sheath 5F, 45 cm, MP, CC H74939277545210 Guiding Sheath 6F, 45 cm, ST, CC H74939277645110 Guiding Sheath 6F, 45 cm, ST, TB H74939277645120 Guiding Sheath 6F, 45 cm, MP, CC H74939277645210 Guiding Sheath 6F, 45 cm, MP, TB H74939277645220 Guiding Sheath 6F, 65 cm, ST, CC H74939277665110 Guiding Sheath 6F, 65 cm, ST, TB H74939277665120 Guiding Sheath 6F, 90 cm, ST, CC H74939277690110 Guiding Sheath 6F, 90 cm, ST, TB H74939277690120 Guiding Sheath 6F, 90 cm, MP, CC H74939277690210 Guiding Sheath 6F, 90 cm, MP, TB H74939277690220 Guiding Sheath 7F, 45 cm, ST, CC H74939277745110 Guiding Sheath 7F, 45 cm, ST, TB H74939277745120 Guiding Sheath 7F, 45 cm, MP, CC H74939277745210 Guiding Sheath 7F, 45 cm, MP, TB H74939277745220 Guiding Sheath 7F, 65 cm, ST, CC H74939277765110 Guiding Sheath 7F, 65 cm, ST, TB H74939277765120 Guiding Sheath 7F, 90 cm, ST, CC H74939277790110 Guiding Sheath 7F, 90 cm, ST, TB H74939277790120 Guiding Sheath 7F, 90 cm, MP, CC H74939277790210 Guiding Sheath 7F, 90 cm, MP, TB H74939277790220 Guiding Sheath 8F, 45 cm, ST, CC H74939277845110 Guiding Sheath 8F, 45 cm, ST, TB H74939277845120 Guiding Sheath 8F, 45 cm, MP, CC H74939277845210 Guiding Sheath 8F, 65 cm, ST, CC H74939277865110 Guiding Sheath 8F, 65 cm, ST, TB H74939277865120 Guiding Sheath 8F, 90 cm, ST, CC H74939277890110 Guiding Sheath 8F, 90 cm, ST, TB H74939277890120 Product Usage: The Chariot Guiding Sheath is designed to perform as an introducer sheath for delivering interventional and diagnostic devices into the peripheral vasculature. The guiding sheath has a coiled shaft design and comes with a straight or preformed multipurpose tip shape. It is equipped with a cross-cut hemostatic valve or Tuohy-Borst adapter to prevent bleeding and a sidearm with a three-way stopcock to allow for flushing and introduction of contrast medium. It is packaged with a dilator to facilitate delivery over a guidewire. The guiding sheath can accommodate guidewires with diameters less than or equal to 0.038 in (0.97 mm). The outer surface of the guiding sheath has a hydrophilic coating from the distal tip to approximately 9 cm from the hub. The distal tip has a radiopaque marker band approximately 6 mm from the distal edge, to help with guiding sheath placement.

Quantity: 7985

Why Was This Recalled?

Boston Scientific is voluntarily recalling its Chariot" Guiding Sheath. To date, Boston Scientific has received fourteen complaints for shaft separation, four for distal shaft separation. The most severe outcome of this failure is embolism of device fragments.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Boston Scientific Corporation

Boston Scientific Corporation has 601 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report