Class III — Low Risk

Low risk — use of or exposure to this product is not likely to cause adverse health consequences.

Blazer Open-Irrigated Ablation Batch Catheter Product Usage: Open Irrigated ablation Recalled by Boston Scientific Corporation Due to Supplier had changed its resin sub-suppliers three (3)...

Date: November 28, 2018
Company: Boston Scientific Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Boston Scientific Corporation directly.

Affected Products

Blazer Open-Irrigated Ablation Batch Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system

Quantity: 1,282 (all three types in Japan)

Why Was This Recalled?

Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.

Where Was This Sold?

Japan