THERMOCOOL SMARTTOUCH SF Bi-Directional Product Codes D-1348-01-S Recalled by Biosense Webster, Inc. Due to Biosense Webster, Inc. has recently received an increased...

Date: September 6, 2017
Company: Biosense Webster, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Biosense Webster, Inc. directly.

Affected Products

THERMOCOOL SMARTTOUCH SF Bi-Directional Product Codes D-1348-01-S, D-1348-04-S, D-1348-05-S

Quantity: 1962 units

Why Was This Recalled?

Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.