SMARTABLATE RF System Generator Recalled by Biosense Webster, Inc. Due to Generator may inadvertently deliver RF energy if the...

Date: June 14, 2017
Company: Biosense Webster, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Biosense Webster, Inc. directly.

Affected Products

SMARTABLATE RF System Generator

Quantity: 185 units total

Why Was This Recalled?

Generator may inadvertently deliver RF energy if the foot pedal is placed in close proximity and oriented toward the Stereotaxis magnets.

Where Was This Sold?

This product was distributed to 15 states: CA, CO, GA, IL, IN, MI, NM, NY, OH, PA, TN, TX, UT, WA, WI