Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

D-10" Hemoglobin Testing System Recalled by Bio-Rad Laboratories, Inc. Due to On a rare occasion the D-10 software might...

Date: February 27, 2014
Company: Bio-Rad Laboratories, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Bio-Rad Laboratories, Inc. directly.

Affected Products

D-10" Hemoglobin Testing System, Model number: 220-0220 Manufactured by: Bio-Rad Laboratories, Inc. 4000 Alfred Nobel Drive Hercules, CA 94545 The D-10 Hemoglobin Testing System is a single module designed for hemoglobin analysis in mid-volume laboratories. The D-10 is a fully integrated system for sample dilution, processing and analysis of hemoglobin, designed for use with specific Bio-Rad reagent kits. The D-10 incorporates the use of a dedicated software system for instrument control, data collection and analysis.

Quantity: 483 instruments in US. OUS to be determined.

Why Was This Recalled?

On a rare occasion the D-10 software might generate an extra result by duplicating the result form the previous sample and the Sample ID/Injection number from the next sample.

Where Was This Sold?

This product was distributed nationwide across all 50 states.