Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

BIO-RAD BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Recalled by Bio-Rad Laboratories, Inc. Due to Due to two issues: 1) Customer complaints associated...

Date: September 17, 2021
Company: Bio-Rad Laboratories, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Bio-Rad Laboratories, Inc. directly.

Affected Products

BIO-RAD BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set CAL The BioPlex 2200 Syphilis Total & RPR Calibrator Set is intended for the calibration of the BioPlex 2200 Syphilis Total & RPR Reagent Pack.

Quantity: 791 units

Why Was This Recalled?

Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.

Where Was This Sold?

This product was distributed nationwide across all 50 states.